Page 3 of 5 APPLICATION NOTE 4.3-10 Calibration Configuration 2 DUT type: 4.3-10 male Test port / cable: N female STEP 1 Calibrate with measuring adapter VNA with N (f) cable Adapter – N (m) to N (f) 33_N-50-0-51 OSL/ECal STEP 2 Replace the 33_N-50-0-51/193 NE adapter with matching 4.3-10 adapter and connect with DUT VNA with N (f) cable Adapter – N (m) to 4.3-10 (f) 33_N-4310-50-1 DUT with 4.3-10 (m) NOTE: For VNA measurement cable with PC 3.5 (f) instead of N (f), replace the 33_N-50-51 adapter with matching 33_PC35-50-0-2 for calibration and 33_PC35-4310-50-1 for measurement. 3.2 METHOD 2: DUT measurement with de-embedding adapters Since calibration kits do not have the same precision and accuracy with 4.3-10 interface as they do with N or PC 3.5 interface, another method to achieve traceable measurements is the use of 4.3-10 de-embedding adapters. With de-embedding adapters, comparable precision to calibration kits can be achieved at a fraction of the cost. First a calibration is done at the reference plane of the VNA cable. To measure a DUT with 4.3-10 interface, a HUBER+SUHNER AG de-embedding adapter is used between VNA cable and the DUT. The influence of the adapter can be considered by loading the S-Parameter of the de-embedding adapters into the NVA. The S-Parameter (supplied by H+S AG via E-Mail, CD or USB device) is then mathematically removed from the DUT measurements. This method is explained in the following example. HUBER+SUHNER disclaims any liability resulting from incorrect installation and use, including any damages resulting from the use of tools and accessories other than the ones recommended herein.