Page 4 of 5 APPLICATION NOTE 4.3-10 Calibration STEP 1 Calibrate your VNA using your preferred method, e.g. OSLT, TRL, or ECal. This may look like Figure 2. Reference plane VNA with N / PC 3.5 cable OSLT / TRL / ECal calibration Figure 2: Calibration with ECal STEP 2 Connect the 4.3-10 adapters and load the H+S supplied data of 4.3-10 adapters into your VNA. Reference plane VNA with N (m) cable HUBER+SUHNER 4.3-10 de-embedding adapter DUT On VNA, this is done by choosing ‘2 port de-embedding’ from the ‘calibration’ menu. Click on the button and load your *.s2p file. Depending on your equipment, the following menu could be different. Figure 3: Load HUBER+SUHNER calibrated adapter data from USB into the VNA and select de-embedding option HUBER+SUHNER disclaims any liability resulting from incorrect installation and use, including any damages resulting from the use of tools and accessories other than the ones recommended herein.